Study for the DANB RHS exam in proportion to the outline. Purpose and Technique alone is 50% of it, and Radiation Characteristics and Protection and Infection Prevention and Control take 25% each. Each section below covers what the questions ask, where answers go wrong, and sample items.
Before you open a textbook
The outline effective March 12, 2025 tests digital radiography only: sensors, phosphor plates, panoramic units and cone beam CT. If your review book has chapters on developer, fixer, safelights or film duplication, skip them; that material left the exam with the move to digital-only content.
Weight tells you where the time goes. With 75 questions, half of them sit in the first domain below, and much of that domain is about images: what you see on them, why one was taken, and what went wrong when it is unusable.
Share of the radiation health and safety outline taken by each of its three domains.
Purpose and TechniqueAnatomy and conditions on images; purpose of each image type; patient preparation and positioning; receptors and holding devices; paralleling, bisecting and panoramic technique; error correction; equipment care; viewing and mounting images; legal requirements for records.
50%
≈38
Radiation Characteristics and ProtectionFactors in x-ray production; types of radiation; radiation biology; units of measure and dose limits; operator and patient protection, including ALARA and image-selection guidelines; equipment malfunction.
25%
≈19
Infection Prevention and ControlStandard precautions for radiographic equipment and supplies under ADA, CDC and OSHA guidance: room setup and breakdown, barriers, clinical contact surfaces, item classification, hand hygiene and PPE.
25%
≈19
Weights: DANB: RHS Exam Outline, effective March 12, 2025 [1]. Read Nov 6, 2026.
Purpose and Technique (50%)
Expect three kinds of question here.
Reading images. Radiopaque structures appear light and radiolucent ones dark; questions ask you to recognize anatomy, restorations, lesions, anomalies and implants on that basis, to tell the arches apart by their landmarks, and to name primary teeth with the Universal letters A to T.
Choosing and taking the image. Each type has a job: bitewings, periapicals, occlusal views, panoramic and cephalometric images, and CBCT. Scenarios describe a clinical question and ask which image answers it, so learn what each one shows and what it cannot. Paralleling keeps receptor and tooth parallel with the beam perpendicular to both; bisecting directs the beam perpendicular to the line that halves the angle between them.
Fixing what went wrong, plus the less glamorous items: caring for sensors and plates, mounting and viewing images, and record rules such as privacy, retention and who owns the images.
Angulation errors come in pairs
Too little vertical angulation stretches the teeth (elongation); too much shrinks them (foreshortening). Wrong horizontal angulation overlaps the contacts; a beam that misses part of the receptor leaves an unexposed area (cone cut). Panoramic errors follow head position in the same paired way: tipping the chin or moving the patient forward or back distorts the occlusal plane and the width of the front teeth, each in its own direction.
Radiation Characteristics and Protection (25%)
This domain is physics, biology and protection.
Production. Questions separate the settings that change the energy of the beam from those that change how many photons it carries, and ask what each does to density and contrast. Learn each setting by its effect on the image, not only by its definition.
Biology. Rapidly dividing cells are the most sensitive. Deterministic effects have a dose threshold and grow more severe with dose; stochastic effects have no threshold, and dose raises only their probability, which is why every exposure has to be justified.
Units. Learn the pairs: gray and rad for absorbed dose, sievert and rem for equivalent and effective dose, coulombs per kilogram and roentgen for exposure in air. Practice converting between the traditional and SI units, including the milli- prefixes on dosimeter reports.
Protection. ALARA, image selection by clinical need, collimation, the position-indicating device, operator distance and barriers, and quality assurance records. Know what to do when the unit malfunctions: stop, report, and keep it out of use until it is checked.
Right number, wrong unit
Answer choices pair a correct figure with the wrong unit: rad beside sievert, rem beside gray. Check that the unit matches the quantity asked for (absorbed dose, equivalent dose or exposure) before you check the number.
Classification. A sensor or phosphor plate touches mucosa, so it is semicritical, yet it cannot go through a heat sterilizer. That combination is behind the receptor questions: what protects them during use and what happens to them between patients.
Setup and breakdown. The order of work matters: what gets covered before the patient sits, what a gloved hand may touch during the series, and how contaminated items leave the room.
Plates. Open the barrier envelope without letting the contaminated outside touch the plate, and process the plate with clean hands.
A barrier is not a substitute for disinfection
A sleeve shields the sensor only while it is in the mouth; afterwards the sensor still needs attention before the next patient. The correct answer pairs the barrier with cleaning and disinfection once it comes off, and treats a visibly soiled barrier as a contaminated surface underneath.
Sample questions from all three domains
0 of 9 done · 0 right
Technique · 50%1 / 9
On images of an 8-year-old with mixed dentition, which finding indicates normal development?
Pick the option you would mark on test day. The choice stays fixed, and every option then explains itself.
No primary teeth remaining in either arch
Primary molars and canines are normally still present at 8; their absence would point to early loss, not normal development.
Permanent tooth buds below primary roots
In mixed dentition the developing permanent teeth sit beneath the roots of the primary teeth they will replace, which is the expected picture at age 8.
Fully formed third molar crowns
Third molar crowns only begin to calcify around ages 8 to 10 and finish years later, so complete crowns at 8 would be unusual.
Complete root formation on all permanent teeth
Complete root formation on all permanent teeth comes in the teen years; at 8 even the first molars and incisors are still finishing their roots.
Technique · 50%2 / 9
Why must imaging software keep each digital image exactly as it was captured?
Edited copies lose the patient ID
Edited copies keep the patient data attached in imaging software, so losing the ID is not the reason.
PSP plates are erased from the saved file
PSP plates are erased by bright light in the scanner after they are read, independent of the saved file.
Unaltered files form the legal record
Unaltered files form the legal record of the examination; enhanced copies are fine for viewing, but the captured image itself must not be overwritten.
Enhanced copies print poorly on paper
Enhanced copies print and share like any other image; print quality is not why the captured version is kept.
Technique · 50%3 / 9
After a maxillary molar extraction, a root tip may be displaced into the sinus. Which image best localizes it in three dimensions?
Repeat periapical of the extraction site
A repeat periapical is a two-dimensional view; it can show the fragment but not its depth relative to the sinus floor.
Panoramic image of both jaws
A panoramic image is two-dimensional and distorted in the sinus region, so it does not show the fragment's depth relative to the sinus floor.
Maxillary topographic occlusal image
An occlusal view adds a different angle but is still a flat projection, not a three-dimensional localization.
Limited-field CBCT of the region
A limited-field CBCT shows the root tip in three planes, so its position relative to the sinus floor can be located with the smallest field that answers the question.
Radiation and protection · 25%4 / 9
With a tungsten target, characteristic radiation from the K shell is produced only when the operating voltage exceeds approximately:
70 kVp
Tungsten's K-shell electrons are bound at about 69.5 keV, so K-shell characteristic radiation appears only once the tube runs at roughly 70 kVp or higher.
50 kVp
At 50 kVp the electrons lack the energy to eject tungsten K-shell electrons, so the useful beam is braking (bremsstrahlung) radiation.
90 kVp
90 kVp is above the threshold, but characteristic radiation already begins near 70 kVp, so 90 is not the point where it starts.
100 kVp
100 kVp is far above the K-shell binding energy; choosing it confuses the usual upper range of dental units with the threshold.
Radiation and protection · 25%5 / 9
Radiation is detected coming through the tubehead housing rather than out of the PID opening. Which type of radiation is this?
Scatter radiation
Scatter radiation is deflected from its path when the beam strikes the patient's tissues; it does not come from inside the housing.
Primary radiation
Primary radiation is the useful beam that exits through the PID opening, not radiation passing through the housing walls.
Leakage radiation
Leakage radiation is any radiation escaping through the tubehead housing other than the useful beam; 21 CFR 1020.30(k) caps it at 0.88 mGy air kerma in 1 hour at 1 meter, and a reading above that means the unit needs service.
Secondary radiation
Secondary radiation is created when the primary beam interacts with matter such as the patient's tissues, not inside the tubehead.
Infection control · 25%6 / 9
While a digital series is being exposed, which practice best keeps the computer free of contamination?
Clicking the mouse with contaminated gloves, then wiping it at noon
Touching the mouse with gloves from the mouth contaminates it for every patient seen before the midday wipe.
Touching the keyboard only with the gloved little finger
Touching the keyboard with the gloved little finger still carries saliva from the mouth onto the keys.
Running the software with a foot pedal or a clean-handed helper
A foot pedal or a helper with clean hands keeps gloves that touched the mouth away from the mouse and keyboard altogether.
Disinfecting the mouse and keyboard once at the end of the day
A single end-of-day disinfection leaves the computer contaminated between patients all day.
Infection control · 25%7 / 9
When imaging a bedridden patient in a nursing facility with a handheld unit, infection control precautions should be:
Identical to those used in the dental office
Identical precautions apply because standard precautions cover every patient in every setting: a bedside handheld exam uses the same barriers, gloves and disinfection as the office.
Limited to hand hygiene between residents
Hand hygiene is necessary but leaves the sensor, holder and handheld unit unprotected between residents.
Lighter than in the dental office
Lighter precautions than in the dental office ignore that saliva and contaminated receptors spread infection the same way at a bedside.
Left to facility housekeeping staff
Leaving it to facility housekeeping staff covers the room, but the radiographer is responsible for the barriers and equipment used on the patient.
Infection control · 25%8 / 9
Placing a latex finger cot over a digital sensor before its barrier sleeve has been shown to:
Eliminate the need for disinfecting the sensor
Sleeves can fail, so the sensor is still cleaned and disinfected after each patient even with a finger cot underneath.
Lower the exposure time the sensor needs
A thin latex layer has no effect on the exposure the sensor requires.
Improve the contrast of the digital image
Contrast depends on exposure settings and image processing, not on the barrier around the sensor.
Reduce leakage of saliva onto the sensor
A finger cot under the sleeve adds a second layer, which has been shown to reduce saliva reaching the sensor through tears or the opening of the sleeve.
Radiation and protection · 25%9 / 9
The quantity of x-rays in the beam is controlled mainly by:
Filtration and collimation
Filtration removes weak photons and collimation trims the beam size; neither is the main control of how many photons are produced.
Milliamperage and exposure time
Milliamperage sets how many electrons cross the tube per second and time sets for how long, so together they determine the number of x-ray photons.
Collimation and PID length
Collimation and PID length shape and direct the beam; they do not set how many photons the tube produces.
Kilovoltage peak and filtration
Kilovoltage peak mainly sets photon energy, and filtration removes weak photons; neither is the main control of beam quantity.
RHS pages
DANB RHSThe dental x-ray exam: what the certificate is, who can sit it and where DANB lists it
Which part of the RHS outline deserves the most study time?
Purpose and Technique, at 50%. The outline lists both image interpretation and error correction inside it.
Do I need to study film processing for the RHS?
No. The exam covers digital radiography only. Film, darkroom and processing chapters in older books can be skipped.
How much does the RHS overlap with the Infection Control exam?
Only where imaging is involved. The radiography exam asks about barriers on sensors and x-ray equipment; sterilization, waterlines and the wider operatory belong to ICE, part of the DANB CDA.